Dead Weight Testers

DWT Series Hydraulic Dead Weight Testers

0.6 bar to 2500 bar · class 0.05, 0.02 or 0.01 · seven models

The standard everything else is checked against. A known mass on a piston of known area produces a known pressure, and nothing else is measured to get there, which is why a dead weight tester is a primary standard and a gauge is not. Seven models, each named by its upper limit in bar, each with a tungsten carbide piston and cylinder and a weight set calibrated to it.

Two M20 × 1.5 output ports on every model · pre-pressurising on the DWT-1000 and DWT-1600 · digital piston position on the DWT-2500 · supplied calibrated

GPS LTD DWT series hydraulic dead weight tester, standard construction: tungsten carbide piston column, two output ports, oil cup and hand pump on a levelled base

Standard construction · DWT-6 to DWT-600

7
Models · 6 to 2500 bar
0.01%
Best Accuracy Class
2500 bar
Top of the Range
Carbide
Piston and Cylinder

Why it is the standard

A gauge measures pressure; a dead weight tester defines it. The number at the output port comes from mass, length and time and nothing in between, so there is no sensor to drift and no electronics to age. Three things make that true.

A known mass

Each weight is calibrated against its own piston and marked with the pressure it produces. The plate and piston alone produce the lower limit of the range.

A known area

The piston and cylinder are tungsten carbide, ground to a measured effective area that does not drift with wear or pressure. That area is the instrument.

A floating piston

Pump until the spinning piston floats between the reference lines. Force over area is then the pressure at the output port, and nothing has been measured to get there.

Which model?

Choose by the highest pressure you need to generate, then the accuracy class by what you calibrate: class 0.05 covers most industrial gauges and transmitters, class 0.01 is for laboratory reference work. The model number is the upper limit in bar.

ModelRangeIn MPaPiston areaWeights produceConstructionWorking liquid
DWT-60.6 to 6 bar0.06 to 0.6 MPa1 cm²0.1 and 0.5 barStandardMixed or transformer oil
DWT-601 to 60 bar0.1 to 6 MPa0.5 cm²1 and 5 barStandardMixed or transformer oil
DWT-2505 to 250 bar0.5 to 25 MPa0.1 cm²5 and 25 barStandardSebacate
DWT-60010 to 600 bar1 to 60 MPa0.05 cm²10 and 50 barStandardSebacate
DWT-100010 to 1000 bar1 to 100 MPa0.05 cm²10, 50 and 100 barPre-pressurisingSebacate
DWT-160020 to 1600 bar2 to 160 MPa0.05 cm²10, 50 and 100 barPre-pressurisingSebacate
DWT-250050 to 2500 bar5 to 250 MPa0.025 cm²10, 20, 100 and 200 barDigital positionSebacate
Each weight is marked in bar for its own instrument. The plate and piston alone produce the lower limit of the range; the weights add their marked pressures on top. Pressure is set by adding weights until the marked values sum to the point required, and the same weight on another instrument produces a different pressure.

Standard construction

DWT-6, DWT-60, DWT-250 and DWT-600. Hand pump, oil cup, two output ports and the measuring column on a levelled base. Reference lines on the column set the working position.

Pre-pressurising system

DWT-1000 and DWT-1600. A pre-pressure valve and shut-off valve fill the system and get it to pressure quickly, so the hand pump does the fine work rather than all of it.

Digital piston position

DWT-2500. An indicator reads the piston height in millimetres, replacing the reference lines. The working band is 0.5 to 2.5 mm and the manual gives the settling routine.

Technical specifications

Common to all seven models except where a model is named. Ranges, piston areas and weight sets are in the model table above; full figures are in the datasheet and the operating manual.
SpecificationDWT series
Accuracy class±0.05%, ±0.02% or ±0.01% of reading, as ordered
Reference conditions20 ±2 °C, temperature drift not exceeding 0.5 °C during a measurement
Piston and cylinderTungsten carbide, precision ground
Plate and piston mass0.6 kg (DWT-6) · 0.51 kg (DWT-60 to DWT-1000) · 1.02 kg (DWT-1600) · 1.2 kg (DWT-2500)
Weight setsDWT-6: 0.102 kg × 6, 0.51 kg × 10 · DWT-60: 0.51 kg × 4, 2.55 kg × 11 · DWT-250: 0.51 kg × 4, 2.55 kg × 9 · DWT-600: 0.51 kg × 4, 2.55 kg × 11 · DWT-1000: 0.51 kg × 4, 2.55 kg × 11, 5.1 kg × 4 · DWT-1600: 0.51 kg × 4, 2.55 kg × 11, 5.1 kg × 10 · DWT-2500: 0.255 kg × 3, 0.5 kg × 4, 2.5 kg × 6, 5.1 kg × 9
Working liquidDWT-6 and DWT-60: mixed or transformer oil, 9 to 12 cSt at 20 °C · DWT-250 and above: diisooctyl sebacate, 20 to 25 cSt at 20 °C · acid value 0.05 mg KOH/g or less
Output connectionsTwo, M20 × 1.5, with union nuts · adapters to other threads on request
Level compensationBy mass or by height · 0.9 g per cm on the DWT-6, 0.45 g per cm on the DWT-60 · union-nut instruments compensated at calibration
Working positionReference lines on the column · DWT-2500: digital indicator, working band 0.5 to 2.5 mm
Piston rotationClockwise, 30 to 60 turns a minute, both hands
Operating environment20 ±5 °C, air free of corrosive gases
Storage5 to 35 °C, relative humidity 80% or less
SparesConnection O-ring 12 × 2.4 mm · pump seal 17 mm cup with 17 × 2.4 mm O-ring to 600 bar, 9.5 (ID) × 2.4 mm at 1000 bar · DWT-1600 and DWT-2500 on request
CalibrationSupplied with a calibration certificate for its class · recalibrate after one year, then at two-year intervals if stable
Warranty1 year, subject to product registration with GPS LTD

Using it well

At class 0.05 and better, the details around the instrument matter as much as the instrument. Three that the operating manual covers in full.

Level compensation

The pressure is at the bottom of the piston cylinder; a sensor higher or lower sees a head of liquid on top. Compensate by mass on the plate or by mounting height. The manual gives the figures.

Working liquid

Mixed or transformer oil on the DWT-6 and DWT-60; diisooctyl sebacate from 250 bar upward for sensitivity at pressure. Filtered, clean, and replaced after a period of use.

A matched set

Piston, cylinder, plate and weights are calibrated together and share a serial number. They are not interchangeable between instruments, even of the same model.

Where they are used

Calibration laboratories

A primary standard on the bench, with class 0.01 available for reference work and traceable certification from GPS.

Transmitters and sensors

Union-nut instruments sit 1.6 cm above the piston base and are compensated for at calibration; connect and go.

Precision pointer gauges

Dial-centre height is the level difference to compensate; the manual works the example for a 150 mm dial.

Plant instrument workshops

Class 0.05 covers most industrial gauges and transmitters at a fraction of the cost of a laboratory standard.

High pressure hydraulics

The DWT-1600 and DWT-2500 reach where general purpose calibrators stop, with pre-pressurising to get there.

Checking other calibrators

A dead weight tester is what a pressure calibrator is checked against. Keep one and the rest of the kit has a reference.

What’s in the box

  • Dead weight tester, model as ordered
  • Matched weight set in its case
  • Two M20 × 1.5 output connections
  • Working liquid for first fill
  • Dust cover
  • Operating manual and calibration certificate

Before you order

Tell us what the instrument will calibrate: the level compensation and the working liquid depend on it, and adapters for connection threads other than M20 × 1.5 are made to order. The instrument works at 20 ±5 °C and the accuracy class applies at 20 ±2 °C, so it belongs on a bench in a room that holds its temperature, not in the plant. The weights are heavy, up to 5.1 kg each, and the largest sets weigh over 60 kg; allow for that in the bench and in who lifts them. The piston, cylinder, plate and weights are a matched set and cannot be replaced individually without recalibration.

Calibration and downloads

Accuracy classes and certification

Class 0.05 Class 0.02 Class 0.01 Calibrated as a set
Certificate
Supplied with every instrument for the class ordered; traceable certification available
Interval
Recalibrate after one year, then at two-year intervals if the first recalibration shows the instrument stable
Service
Calibration, repair and seal kits from GPS; the piston and cylinder are not user-serviceable
Warranty
1 year, subject to product registration with GPS LTD

Request a quote — DWT Series

We’ll reply with pricing, lead time and options.

Calibration — DWT Series

Tell us the model and class and we’ll confirm the certificate options and cost.

General enquiry

Ask us anything — sales, technical, or order related.

Request a Quote

We’ll reply with pricing, lead time and options.