Questions we get asked
Gauge, absolute or differential: which do I need?
Gauge pressure is measured against ambient, absolute against a vacuum, and differential between two ports. Choose the one your process uses as its reference; a gauge-referenced instrument on an absolute process reads wrong by one atmosphere and looks fine.
How much accuracy should I specify?
As a rule, four to ten times better than the tolerance you are checking. A 0.5 percent process gauge wants a 0.05 percent reference. For calibration work, look at the stability over a year and the temperature effect across your range as well as the headline figure.
Hydraulic or pneumatic test pump?
Pneumatic for vacuum and low pressure air, hydraulic for anything higher. Check the media the instrument under test can take, oil or water, and the fine-adjust resolution you need to sit on a calibration point.
When do I need a dead weight tester rather than a calibrator?
When the instrument under test is itself a reference, or when you need the uncertainty to come from mass and area rather than from a sensor that drifts. A dead weight tester defines the pressure; a calibrator measures it. Most workshops use a calibrator day to day and a dead weight tester to check the calibrator.