1.0 to 1000 V DC · 100 kΩ to 10 TΩ · up to 55 readings/s
Insulation is judged at the voltage it has to stand, so the GPS-3861C lets you choose that voltage to a tenth of a volt anywhere from 1 V to 1000 V, then reads the resistance and the leakage current side by side. From a capacitor on a sorting line to a reel of cable on the bench.
Charge timer to 999.9 s · open circuit zero on every range · PASS, UPPER and LOWER sorting · Handler, RS-232 with SCPI and USB logging

At gigaohms and teraohms the instrument is reading picoamps, and the way the test is set up matters as much as the meter. All three of these are under your control.
Pick one of the common voltages from a soft key, 50, 100, 250, 500 or 1000 V, or type any value in 0.1 V steps. The six ranges move with it: at 100 V the meter reads 1 MΩ to 1 TΩ, and at 1000 V the top range reaches 10 TΩ.
A cable or a capacitor draws charging current before it settles, and a reading taken too early is low. The charge timer holds the voltage on for up to 999.9 s before the reading is taken, and the contact check catches a capacitive part that is not actually connected.
Surface leakage across a dirty or damp insulator flows in parallel with the resistance you want. Taking the shield or guard to the ground terminal steers that current around the measurement, and an open circuit zero stored for every range removes the leads’ own contribution.
Accuracy at the slow rate: ±5% below 1 MΩ, ±1% above 1 MΩ, ±3% above 1 GΩ, ±5% above 10 GΩ and ±10% above 1 TΩ, at 23 °C ±5 °C and 65% relative humidity, after an open circuit zero and more than 60 minutes warm-up.
Every measurement runs the same three states, and the instrument shows which one it is in. That is what makes a high-voltage test safe to repeat a hundred times.
The test voltage goes on and the charge timer starts. Skip it with the charge time at zero, or press TEST to go straight to the reading.
Resistance, leakage current and the voltage at the terminals are read together, with the sorting result when the comparator is on.
The voltage comes off and the part is discharged quickly through the instrument. With automatic discharge on, this follows every reading by itself.
The device under test is connected and removed only in the discharge state. Polarised parts such as electrolytic capacitors go positive to (+) and negative to (−).
Six ranges, chosen automatically, held, or set from the comparator limit, cover a span that slides upwards as the test voltage rises. These are the spans at eleven typical voltages.
| Voltage | Range 1 | Range 2 | Range 3 | Range 4 | Range 5 | Range 6 |
|---|---|---|---|---|---|---|
| 1 V | Not specified | 1M – 10M | 10M – 100M | 100M – 1G | 1G – 10G | 10G – 100G |
| 10 V | 100k – 1M | 1M – 10M | 10M – 100M | 100M – 1G | 1G – 10G | 10G – 100G |
| 25 V | 250k – 2.5M | 2.5M – 25M | 25M – 250M | 250M – 2.5G | 2.5G – 25G | 25G – 250G |
| 50 V | 500k – 5M | 5M – 50M | 50M – 500M | 500M – 5G | 5G – 50G | 50G – 500G |
| 75 V | 750k – 7.5M | 7.5M – 75M | 75M – 750M | 750M – 7.5G | 7.5G – 75G | 75G – 750G |
| 100 V | 1M – 10M | 10M – 100M | 100M – 1G | 1G – 10G | 10G – 100G | 100G – 1T |
| 125 V | 2.5M – 12.5M | 12.5M – 125M | 125M – 1.25G | 1.25G – 12.5G | 12.5G – 125G | 125G – 1.25T |
| 250 V | 5M – 25M | 25M – 250M | 250M – 2.5G | 2.5G – 25G | 25G – 250G | 250G – 2.5T |
| 500 V | 10M – 50M | 50M – 500M | 500M – 5G | 5G – 50G | 50G – 500G | 500G – 5T |
| 750 V | 10M – 75M | 75M – 750M | 750M – 7.5G | 7.5G – 75G | 75G – 750G | 750G – 7.5T |
| 1000 V | 10M – 100M | 100M – 1G | 1G – 10G | 10G – 100G | 100G – 1T | 1T – 10T |
Units are ohms: k = 10³, M = 10⁶, G = 10⁹, T = 10¹². The lowest range is not specified at 1 V. For leakage current measurements a held range is recommended.
Slow for the bench, fast for a machine. The same instrument does both, and the sorting decision goes straight to the PLC.
The Handler outputs pass, fail, open circuit and end of test, with a pair of relay contacts for the sorting result, and takes trigger, charge, discharge and keypad lock inputs. Everything is optocoupler isolated and runs from your own 3.3 to 24 V DC supply. The charge and discharge inputs will also take a foot switch.
RS-232 carries a full SCPI command set at up to 115200 baud, with results sent on request or automatically after every test. Readings can also be logged straight to a USB flash drive as CSV files, on a timer, on every trigger or at the press of a key.
| Measures | Insulation resistance and leakage current, shown with the test voltage |
|---|---|
| Resistance range | 100 kΩ to 10 TΩ across six ranges · maximum reading 9999 |
| Accuracy | ±5% below 1 MΩ · ±1% above 1 MΩ · ±3% above 1 GΩ · ±5% above 10 GΩ · ±10% above 1 TΩ |
| Test voltage | −1.0 to −1000 V DC in 0.1 V steps · accuracy ±10% below 10 V, ±1% above · maximum charging current 30 mA ±5 mA |
| Measurement rate | Slow about 3, medium about 25, fast about 55 readings per second |
| Ranging | Auto, hold, or nominal from the comparator lower limit |
| Charge timer | 0 to 999.9 s · accuracy ±0.5% · counts up or down · automatic discharge on or off |
| Correction | Open circuit zero for all ranges in one operation · contact check for capacitive parts above 100 pF |
| Comparator | PASS, UPPER and LOWER · beeper off, on good or on fail |
| Trigger | Internal, manual, external on a rising or falling edge, or bus · delay 1 ms to 60 s |
| Display | 3.5 in true-colour TFT · English interface · keypad lock |
| Interfaces | RS-232 with SCPI, 1200 to 115200 baud · Handler · USB flash drive, CSV logging · RS485 and USB-to-RS-232 optional |
| Handler | Pass, fail, open circuit and end of test outputs · relay contacts · trigger, charge, discharge and keypad lock inputs · optocoupler isolated · external 3.3 to 24 V DC supply |
| Access | Administrator and user accounts with an optional password |
| Power | 198 to 242 V AC · 50/60 Hz · 30 VA maximum · 250 V 1 A slow-blow fuse |
| Environment | Operating 10 to 40 °C, 10% to 80% RH · storage 0 to 50 °C, 10% to 90% RH |
| Dimensions · weight | 264 × 107 × 350 mm (W × H × D) · about 5 kg |
RS485 and USB-to-RS-232 interfaces are available to order. Tell us in your enquiry if you need either, or if you would like the instrument supplied with a UKAS calibration certificate.

We’ll reply with pricing, lead time and options.
Send us the serial number and we’ll locate your certificate.
Ask us anything — sales, technical, or order related.
We never share your details. By sending, you agree to be contacted about your enquiry.