Electrical Safety Tester

GPS-5030EC Earth Continuity Tester

3 to 30 A · 0 to 600 mΩ · four-terminal Kelvin

A continuity check at a few milliamps will pass a joint that cannot carry fault current. This one forces up to 30 A through the protective earth path and reads the bond in milliohms, then judges it against a window and says so in about a second.

Download Datasheet

0.1 mΩ resolution · upper and lower limit judgement · RS232 and remote control port · part number 861030

GPS LTD GPS-5030EC earth continuity tester, front panel showing the vacuum fluorescent display, four-terminal Kelvin connections and START and STOP controls

GPS-5030EC · 380 × 320 × 130 mm, 10 kg

Why four terminals

A good bond is a few tens of milliohms. So is a metre of test lead. Measure with two wires and you are largely measuring your own leads, which is why this instrument uses four and prints the diagram on its front panel.

OUTPUT carries the current, SAMPLING senses across the bond
The current pair
OUTPUT drives up to 30 A through the earth path. Whatever resistance the leads and clips add here does not matter, because the current is held constant regardless.
The sense pair
SAMPLING measures the volt drop across the bond alone. Almost no current flows in these leads, so they contribute almost no error.
What is left over
Whatever the fixture still adds is nulled with the resistance offset before testing starts, so the reading is the product and nothing else.

Choosing the current

Pick the current your standard calls for first, then check the resistance you expect falls under the ceiling for it. The output has to develop the volt drop across the bond, so the higher the current, the lower the resistance it can reach.

30.0 A
Test current
200 mΩ
Highest current, tightest ceiling
25.0 A
Test current
270 mΩ
Maximum resistance
20.0 A
Test current
390 mΩ
Maximum resistance
10.0 A
Test current
600 mΩ
Full range, lowest current

Anywhere from 3.0 to 30.0 A can be set in 0.1 A steps. The four figures above are the published points; between them the ceiling moves the same way. If a reading sits close to the limit for your chosen current, drop the current a step and the headroom returns.

Two limits, three verdicts

Both an upper and a lower limit are set, anywhere across the full 0 to 600 mΩ span. A single threshold would pass a shorted fixture; a window does not.

Pass

Inside the window

The reading sits between the lower and upper limits for the whole test time. The instrument holds the result, sounds the pass tone and lights PASS.

Fail upper

Resistance too high

A poor joint, a corroded contact, a screw that was never tightened, or a bond that simply is not there. The display distinguishes this from a low fail.

Fail lower

Resistance too low

Usually not the product. A shorted fixture, a probe clipped to the wrong point, or leads still shorted together from the offset null.

The display carries UPPER and LOWER as separate indications, so an operator can tell a bad product from a bad connection without reading the number. An open circuit raises its own alarm the instant it is detected.

On the line

Built to be started by something other than a person

This is a station rather than a handheld. The test runs to a set time, returns a verdict on its own, and can be driven from a controller so the operator only presents the product.

Remote port
A front-panel DIN connector takes start and stop from a PLC or a foot switch, so the cycle can be tied to a fixture closing.
RS232
Settings and results over a serial link to a PC or test controller.
Timed cycle
0.5 to 999.9 s in 0.1 s steps, or continuous. The instrument ends the test and judges it without intervention.
Keylock
Locks the front panel so a set-up cannot be changed by accident once the line is running.
Duty cycleThe front panel is marked AC on 30 minutes, off 30 minutes. Worth designing the cycle around before committing the instrument to a continuous line.

Technical specifications

Resistance and current figures are quoted for four-terminal measurement with the fixture nulled by the resistance offset.
Ground resistance0.0 to 600 mΩ · ±(2% rdg + 0.3 mΩ) · 0.1 mΩ resolution · four-terminal Kelvin measurement
Limit settingsUpper and lower, each 0.0 to 600 mΩ · ±(2% set + 0.3 mΩ) · 0.1 mΩ resolution
Test current3.0 to 30.0 A · measured ±(2% rdg + 0.2 A), set ±(2% set + 0.2 A) · 0.1 A resolution · true RMS
OutputSine wave, crest factor 1.3 to 1.5 · distortion below 2% on no load or a pure resistive load · 50 Hz or 60 Hz, selectable
Maximum resistance200 mΩ at 30.0 A · 270 mΩ at 25.0 A · 390 mΩ at 20.0 A · 600 mΩ at 10.0 A
Test timeSet 0.5 to 999.9 s, or 0 for continuous · ±(0.2% set + 0.05 s) · measured ±(0.1% set + 0.05 s) · 0.1 s resolution
Duty cycleAC on 30 minutes, off 30 minutes, as marked on the front panel
JudgementWindow comparator against upper and lower limits · separate UPPER and LOWER fail indications · audible and visual alarm
ProtectionInstant alarm if the earth path is open circuit · PROTECTION annunciator on the display
OffsetAdjustable resistance offset, to null the fixture and leads before testing
DisplayVacuum fluorescent, showing test current, resistance and elapsed time together, with READY, TEST, PASS, FAIL, KEYLOCK and REMOTE annunciators
ControlsFront-panel output control · UP / LOW and SHIFT keys · timer ON/OFF · keylock · START and STOP
InterfacesRS232 · remote control port on a front-panel DIN connector for start and stop from a line controller
Power220 V AC ±10%, single phase, 47 to 63 Hz · fuse 5 A / 250 V AC
EnvironmentOperating 0 to 40 °C · relative humidity 75% or below
Dimensions · weight380 × 320 × 130 mm (l × w × h) · 10 kg
Part number861030

Request a quote — GPS-5030EC

We’ll reply with pricing, lead time and options.

Request a calibration certificate

Send us the serial number and we’ll locate your certificate.

General enquiry

Ask us anything — sales, technical, or order related.

Request a Quote

We’ll reply with pricing, lead time and options.

We never share your details. By sending, you agree to be contacted about your enquiry.