Earth Testing Kit

ERT-50R Double Clamp Multifunction Earth Tester

Staked, one clamp or stakeless · 0.001 Ω · earth current to 600 A

Driving spikes is the accurate way to measure an earth electrode, and it is often the one thing a site will not let you do. A twelve-position switch carries every method in current use, including two that need no disconnection and one that needs no soil at all, so you get a reading indoors, up a pylon, or on a concrete floor.

Download Datasheet

Twelve switch positions · six terminals with jack indicator lamps · 2000-group memory over USB and Bluetooth · part number 800350

GPS LTD ERT-50R double clamp multifunction earth tester, front panel showing the twelve-position rotary switch, backlit display and six-terminal block
ERT-50R · 280 × 260 × 160 mm, 1.80 kg

Three ways in

Every earth resistance method is a trade between what the site will let you do and how good a number you get back. The ERT-50R carries all three, so the site decides, not the instrument.

Staked

Two, three or four wire

The reference measurement. Four-wire takes the leads and the connection out of the result, which on a low-value grid is most of the reading. Nothing else here is this precise.

Needs
Spikes in the ground, electrode disconnected
Gets you
0.001 Ω to 300 kΩ at ±2%
Costs you
You have to break the bond
One clamp

Selection method

A clamp round one leg of a multi-grounded system separates that leg's current from the rest, so you get the resistance of a single electrode while the system stays bonded and live.

Needs
Spikes in the ground, nothing disconnected
Gets you
0.001 Ω to 3000 Ω at ±2%
Costs you
Still needs soil for the spikes
Stakeless

Double clamp

No spikes at all. One clamp injects, the other reads, and the instrument returns the loop resistance. This is what lets you test inside a building, up a pylon, or on a concrete floor.

Needs
Neither. Two clamps and a loop
Gets you
0.01 to 100 Ω at ±10%
Costs you
Coarser, and it needs a multi-grounded return path

All three use the same six terminals. Measure earth voltage before resistance on any of them: a stray potential in the ground corrupts the reading, and above about 5 V of interference the instrument shows NOISE and holds rather than return a wrong figure.

Beyond resistance

The two clamps are not only there to make the stakeless method work, and with six terminals and twelve switch positions the instrument takes some care to stop you wiring a test up wrongly.

The clamps read current too

The same matched pair that drives the double clamp method also measures what is actually flowing in an earth conductor, which is a different question from how good the electrode is. Circulating current in a bonded system, or leakage that should not be there at all.

Range
AC 0.00 mA to 600.0 A, resolving 0.01 mA
Jaw
φ70 mm, 2 m lead, supplied as a matched pair
Marking
MAX 600 V CAT III on the clamps themselves

A lamp at every jack

Six terminals, E, ES, P(S), C(H), C1 and C2, and a method on each of twelve switch positions. Turn the switch and a lamp lights at each jack the selected method uses, showing which lead goes where, so a miswired test is caught before it is run rather than after.

Test current
250 mA short circuit, for driving a reading through poor or stony ground
Measurement
About 7 s, with a countdown on the display
Storage
2000 groups, out over USB or Bluetooth

Technical specifications

Accuracy is quoted at 23 ±5 °C with relative humidity below 75%. Earth resistance ranges are selected automatically.
Earth resistance0.000 to 9.999 Ω · ±(2% rdg + 0.05 Ω) · 0.001 Ω resolution
 10.00 to 29.99 Ω · ±(2% rdg + 5 digits) · 0.01 Ω resolution
 30.0 to 299.9 Ω · ±(2% rdg + 3 digits) · 0.1 Ω resolution
 300 to 2999 Ω · ±(2% rdg + 3 digits) · 1 Ω resolution
 3.00 to 300.00 kΩ · ±(4% rdg + 3 digits) · 10 Ω resolution
Selection method0.000 Ω to 3000 Ω · accuracy as the staked ranges · one clamp with the stakes in place
Double clamp0.01 to 0.99 Ω · ±(10% rdg + 5 digits) · 0.01 Ω resolution
 1.0 to 29.9 Ω · 0.1 Ω resolution · 30 to 100 Ω · 1 Ω resolution
Soil resistivity0.01 Ω·m to 9000 kΩ·m by the Wenner four-pole method · ρ = 2πaR · follows the earth resistance accuracy
Earth voltageAC 0.0 to 750.0 V · DC 0.0 to 1000.0 V · ±(2% rdg + 3 digits) · 0.1 V resolution
Earth currentAC 0.00 mA to 600.0 A · ±(2% rdg + 3 digits) · 0.01 mA resolution · clamp CT, φ70 mm jaw
DC resistanceAvailable on the 2-wire and 4-wire switch positions
Test signalAC 250 mA maximum short circuit, sine wave · open-circuit 40 V AC maximum
Test frequency128 / 111 / 105 / 94 Hz, selected automatically (AFC), with FFT interference analysis
Electrode spacing1 to 100 m, entered on the instrument
Measurement timeAbout 7 s per earth resistance measurement · about two readings per second on earth voltage and current
Auxiliary earth test0.00 kΩ to 30.00 kΩ (100R + rC < 50 kΩ, 100R + rP < 50 kΩ)
Controls12-position rotary switch, one position per method · TEST, HOLD, SET, AL, CLR, READ and backlight keys
TerminalsSix: E, ES, P(S), C(H), C1 and C2, each with a jack indicator lamp showing the correct lead
Display128 × 75 mm LCD, 124 × 67 mm active area, 4 digits, white backlight, bar graph, countdown during measurement
IndicationsHOLD, READ, OL overflow, NOISE interference, FULL memory, low battery
Memory · interfaces2000 groups · USB for upload, saving, analysis and printing · Bluetooth to an Android app
Current clampsTwo matched sensing transformers, CT1 and CT2 · φ70 mm jaw · 2 m lead · marked MAX 600 V CAT III
PowerDC 7.4 V 2600 mAh rechargeable lithium pack · more than 5000 measurements per charge · auto power off after 15 minutes
ProtectionIP65 with the case closed · overload AC 280 V for 3 s across C(H) to E and P(S) to ES
SafetyCAT III 300 V · CAT IV 150 V · IEC 61010-1, pollution degree 2 · IEC 61557-1 and 61557-5 · IEC 61326 · CE
EnvironmentOperating −10 to 40 °C, below 80% RH · storage −20 to 60 °C, below 70% RH
Dimensions · weight280 × 260 × 160 mm · 1.80 kg with battery · about 6.13 kg complete kit
Part number800350

What’s in the kit

ERT-50R kit packed in its carry bag, the instrument in its yellow IP65 case with the four lead reels and the earth stakes stowed alongside
The complete kit, about 6.13 kg packed
  • ERT-50R tester in its own IP65 case, part number 800350
  • Current clamps, two matched sensing transformers, φ70 mm jaw, 2 m lead
  • Standard test leads: 20 m red, 20 m black, 10 m yellow, 10 m green
  • Simple test leads: 1.6 m red, 1.6 m black
  • Auxiliary earth stakes, four, φ9 × 230 mm
  • USB communication cable with PC software
  • Battery charger
  • Instruction manual and tool bag

Everything for all three methods travels together. The four reels and four stakes cover the staked and selection tests and a four-pole Wenner spread; the matched clamp pair covers the stakeless method and current measurement. Nothing needs to be ordered separately to use any switch position.

Request a quote — ERT-50R

We’ll reply with pricing, lead time and options.

Request a calibration certificate

Send us the serial number and we’ll locate your certificate.

General enquiry

Ask us anything — sales, technical, or order related.

Request a Quote

We’ll reply with pricing, lead time and options.

We never share your details. By sending, you agree to be contacted about your enquiry.