2 200 000 count dual display · true RMS · four-wire resistance with offset compensation
Six integration settings, from one reading every two seconds at the full 2 200 000 counts to ten thousand a second at 2.7 ppm of range in noise. The accuracy figures belong to the slow end, and this page says so wherever a rate appears.

Selectable 10 MΩ or 10 GΩ DC input · USB, LAN and SCPI · CAT I 1000 V and CAT II 600 V
This is the setting that decides what the instrument is. Integration time runs from 0.005 to 100 power line cycles, and the reading rate with it, across a range of twenty thousand to one.
Noise below is the source’s own figure for RMS noise as a proportion of range. The published accuracy specifications hold at 100 NPLC; every setting below it adds a stated noise figure on top, so the trade is quantified rather than implied.

Scan down for the function, across for the range. Where a row is all dashes the instrument does not do it, so the table reads the same way on every meter in this category.
| Function | Lowest | Highest | Best resolution | Basic accuracy | Notes |
|---|---|---|---|---|---|
| DC voltage | 200 mV | 1000 V | 0.1 µV | ±(0.0035% + 0.0006) | Five ranges · 10 MΩ, or above 10 GΩ on 200 mV, 2 V and 20 V |
| AC voltage | 200 mV | 750 V | 0.1 µV | ±(0.06% + 0.03) | True RMS · 3 Hz to 300 kHz · figure is 10 Hz to 20 kHz |
| DC current | 200 µA | 10 A | 0.1 nA | ±(0.050% + 0.002) | Six ranges · above 7 A, thirty seconds on and thirty off |
| AC current | 200 µA | 10 A | 0.1 nA | ±(0.10% + 0.04) | Six ranges · true RMS · figure is 10 Hz to 5 kHz |
| Resistance | 200 Ω | 100 MΩ | 0.1 mΩ | ±(0.010% + 0.001) | Seven ranges · offset compensation on the three lowest |
| Capacitance | 2 nF | 100 mF | — | ±(1% + 0.1) | Nine ranges · ±(2% + 2.4) on the 2 nF range |
| Frequency | 3 Hz | 1 MHz | — | ±0.007% | Measured on the 200 mV to 750 V AC ranges · 1 s gate |
| Frequency, high | — | — | — | — | Not a function of this instrument |
| Period | 1 µs | 333.33 ms | — | ±0.007% | Same input ranges as frequency |
| Temperature, RTD | −200 °C | 660 °C | — | 0.16 °C | Platinum, α = 0.00385 · four-wire, or two-wire with REL |
| Temperature, thermocouple | −150 °C | 1820 °C | — | 0.5 °C | Eight types · built-in cold junction, ±3.5 °C |
| dBm | — | — | — | — | Computed from the voltage reading · reference impedance settable |
| Continuity | — | 2 kΩ | — | ±(0.010% + 0.020) | Fixed range · 1 mA test current · threshold 1 Ω to 2 kΩ |
| Diode | — | 4 V | — | ±(0.010% + 0.020) | 1 mA test current · threshold adjustable across the range |
| Rotation speed | — | — | — | — | Not a function of this instrument |
| Square wave out | — | — | — | — | Not a function of this instrument |
Accuracy is specified for one year after calibration, at the calibration temperature ± 5 °C, after ninety minutes of warm-up and at 100 NPLC integration, in the form ±(% of reading + % of range). Each figure is the best available across that function’s ranges, so it does not hold on every range. Resolution is the source’s own published figure. Range-by-range detail is in the datasheet.
Resolution is the easy half. What decides whether the last digit means anything is what sits between the instrument and the part.
Below an ohm or so the leads and the contacts contribute more than the part does, and no amount of resolution fixes that. Two terminals drive the current and two sense the voltage, so the reading is taken at the part rather than at the end of a cable.
Four wires remove the lead resistance but not the thermal EMF at the junctions, which is what is left when a low value measurement still will not settle. On the 200 Ω, 2 kΩ and 20 kΩ ranges the instrument measures with the source on and off and subtracts the difference.
On the 200 mV, 2 V and 20 V ranges the DC input impedance lifts from 10 MΩ to above 10 GΩ. On a high impedance source that is the difference between measuring a voltage and changing it, and here it is a setting rather than a second instrument.
The published figures assume ninety minutes of warm-up, which is a real condition rather than a formality on an instrument resolving to a tenth of a microvolt. A calibration certificate ships with it, so the first reading is traceable rather than merely plausible.
The same instrument, driven rather than read. Nothing below is an option except the GPIB adapter.
Six hundred million samples on one trigger is not a number anyone needs in full, but it is the difference between a capture that is bounded by the instrument and one that is bounded by the job.
Three things that are easier to know now than after the instrument arrives.
The full 2 200 000 counts belong to 100 NPLC, which is one reading every two seconds, and that is where every accuracy figure on this page is specified. Ten thousand readings a second is the other end of the same dial, at 2.7 ppm of range in noise rather than 0.035. Both are real; they are not both available at once.
The panel is marked CAT I 1000 V and CAT II 600 V, which are two permissions rather than one range: up to 1000 V on circuits isolated from the mains, but 600 V once the leads are on a mains-connected circuit. The 750 V AC and 1000 V DC ranges are therefore CAT I only.
A selector on the rear panel sets the instrument to 100 to 120 V or 200 to 240 V. A unit shipped to the UK should already be on the higher band, but it is worth confirming before the first power-up. Line frequency is detected automatically. Consumption is 25 VA maximum.
The terminal block is marked 1000 V max, 500 Vpk max to earth and 10 A rms, fused on the rear panel.
A USB to GPIB adapter, for rigs already on a GPIB bus. The scanner card is not an accessory: it is fitted at build to a separate variant and cannot be added afterwards, so if multi-point switching is part of the job it has to be specified with the order. Everything on the matrix works with what is supplied.
We’ll reply with pricing, lead time and options.
Send us the serial number and we’ll locate your certificate.
Ask us anything — sales, technical, or order related.
We never share your details. By sending, you agree to be contacted about your enquiry.